  2. An automatic monitor of formaldehyde in air by a monitoring tape method

An automatic monitor of formaldehyde in air by a monitoring tape method

  • J Environ Monit. 1999 Jun;1(3):255-8. doi: 10.1039/a900410f.
N Nakano 1 K Nagashima
Affiliations

Affiliation

  • 1 Riken Keiki Co., Ltd., 2-7-6, Azusawa, Itabashi-ku, Tokyo 174-8477, Japan.
Abstract

An automatic monitor has been developed for measuring formaldehyde in air using a sensitive tape for formaldehyde. It is based on the color change of the tape on reaction with formaldehyde. The porous cellulose tape, containing silica gel as an absorbent and impregnated with the processing solution containing hydroxylamine sulfate, Methyl Yellow (pH indicator; pH 2.9-4.0, red-yellow), glycerin and methanol, was found to be a highly sensitive means of detecting formaldehyde and maintains a stable sensitivity. When the tape was exposed to a sample of air containing formaldehyde, the color of the tape changed from yellow to red. The degree of color change was proportional to the concentration of formaldehyde at a constant sampling time and flow rate, and it could be recorded by measuring the intensity of reflected LIGHT (555 nm). The tape could be used to detect down to 0.08 ppm (World Health Organization standard) of formaldehyde with a sampling time of 30 min and a flow rate of 100 mL min-1. Reproducibility tests showed that the relative standard deviation of response (n = 10) was 3.8% for 0.1 ppm formaldehyde. The monitor is simple, specific, capable of unattended operation and is recommended for both laboratory and field operation.